A Study of the Complexation of Mercury(II) with Dicysteinyl Tetrapeptides by Electrospray Ionization Mass Spectrometry
Published on: January 8, 2016
[Determination of mercury by electric heat digestion and atomic fluorescence spectrometry]
Lisha Mao1, Yongru Li, Honghe Liu
1Shenzhen Center for Disease Control and Prevention, Shenzhen 518020, China.
Objective:
To develop a method to determine mercury by electric heat digestion and atomic fluorescence spectrometry.
Methods:
After electric heat digestion using aqua regia, the soil samples were determined by atomic fluorescence spectrometry.
Results:
The linear range of the method was from 0 to 50 microg/L. The correlation coefficient was 0.9994. The detection limit of the method was 0.0051 microg/L. The relative standard deviations ranged from 1.55% to 6.52%. Four soil samples for quality control were determined by this method and all of the results were in the range of the true values.
Conclusion:
The optimized method have many advantages, such as high sensitivity, few interferences, wide linearity. At the same time,it is simple and rapid.
More Related Videos
06:53Preparation of Food Samples Using Homogenization and Microwave-Assisted Wet Acid Digestion for Multi-Element Determination with ICP-MS
Published on: December 22, 2023
08:21Determination of Inorganic Arsenic in a Wide Range of Food Matrices using Hydride Generation - Atomic Absorption Spectrometry.
Published on: September 1, 2017
Related Concept Videos
Atomic Absorption Spectroscopy: Atomization Methods
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...
Atomic Fluorescence Spectroscopy
Gas Chromatography: Types of Detectors-II
Atomic Emission Spectroscopy: Lab
Atomic Emission Spectroscopy: Instrumentation