Biomechanical effect of platform switching in implant dentistry: a three-dimensional finite element analysis
Chih-Ling Chang1, Chen-Sheng Chen, Ming-Lun Hsu
1Department of Dentistry, National Yang-Ming University, Tapei, Taiwan. mlhsu@ym.edu.tw
Purpose:
The purpose of this study was to analyze and compare the implant-bone interface stresses in anisotropic three-dimensional finite element models of an osseointegrated implant with platform switching and a conventional matching-diameter implant platform and abutment in the posterior maxilla.
Materials And Methods:
Three-dimensional finite element models were created of a first molar section of the maxilla and embedded with a single endosseous implant (4.1 3 10 mm). One model simulated a 4.1-mm-diameter abutment connection and the other was a narrower 3.4-mm-diameter abutment connection, ie, simulating a platform-switching configuration. A gold alloy crown with 2-mm occlusal thickness was applied over the titanium abutment. Material properties of compact and cancellous bone were modeled as fully orthotropic and transversely isotropic, respectively. Oblique (200-N vertical and 40-N horizontal) occlusal loads were applied and perfect bonding was assumed at all interfaces.
Results:
Maximum von Mises, compressive, and tensile stresses in compact bone were lower in the platform-switching model than in the conventional model. However, the maximum von Mises stress in cancellous bone was higher in the platform-switching model than in the conventional model.
Conclusion:
The platform-switching technique reduced the stress concentration in the area of compact bone and shifted it to the area of cancellous bone during oblique loading.