Quantitative Detection of Trace Explosive Vapors by Programmed Temperature Desorption Gas Chromatography-Electron Capture Detector
Published on: July 25, 2014
The role of stationary phase selection on performance for explosives analysis using GC-ECD
Roderick R Kunz1, Kerin E Gregory, Dennis Hardy
1Lincoln Laboratory, Massachusetts Institute of Technology, Lexington, MA 02420, USA.
Abstract:
Gas chromatography with electron capture detection (GC-ECD) analysis of explosive-related nitro organic compounds was performed using four different column stationary phases with the focus being on their impact on analyte stability and transfer efficiency during analysis. All four columns used were 6 m x 0.53 mm, and the four stationary phases were a 1.0-microm thick 5% phenyl siloxane/95% methyl siloxane non-polar phase, a 1.5-microm thick 5% phenyl siloxane/95% methyl siloxane non-polar phase optimized for explosives analysis, an intermediate polarity 0.5-microm thick trifluoropropylmethyl siloxane phase, and a proprietary intermediate polarity 0.5-microm thick phase. Although all exhibited similar recovery (as defined as the detector signal per injected mass) when new, the intermediate polarity phases maintained higher sample recovery over the course of analyzing hundreds of samples than the non-polar phases, particularly for the nitramines hexahydro-1,3,5-trinitro-1,3,5-triazine and octahydro-1,3,5,7-tetranitro-1,3,5,7-tetrazocine, for which a 7x and 3x decrease in recovery were observed, and the nitrate esters nitroglycerin and pentaerythritol tetranitrate, for which a 7x and 11x decrease in recovery were observed. For most other explosive-related compounds, the differences in recovery were between 1.5x and 3x over the same course. Although the detailed chemical formulation of the stationary phases have not been disclosed by their manufacturers, we attribute the observed differences in performance to the stability of their passivation chemistries with regard to other mobile-phase compounds contained in complex field samples. Although these effects have been qualitatively noted in the past and in response, maintenance procedures have been developed to help account for this behavior, the analyst's preference is to use an explosives analysis method that does not require these time-consuming measures. Our desire to prolong this maintenance interval provided the motivation for the assessment reported in this paper. From our assessment, we conclude that manufacturers of GC columns should focus more attention on the stationary phase and passivation chemistries that can lead to the development of a column that is better able to maintain passivation against explosive compound degradation; and users intending to perform large numbers of analyses using GC-ECD should make this a consideration when selecting a column.
More Related Videos
05:31Gas Chromatography-Mass Spectrometry Paired with Total Vaporization Solid-Phase Microextraction as a Forensic Tool
Published on: May 25, 2021
08:56Detection of Regulated Ergot Alkaloids in Food Matrices by Liquid Chromatography-Trapped Ion Mobility Spectrometry-Time-of-Flight Mass Spectrometry
Published on: November 22, 2024
Related Concept Videos
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...