Objectives:
The aim of this study was to evaluate an algorithm that estimates changes in cardiac output (CO) from right ventricular (RV) pressure waveforms derived from an implantable hemodynamic monitor (IHM) in heart failure patients.
Design:
Twelve heart failure patients (NYHA II-III, EF 32%) with an implantable hemodynamic monitor (Chronicle) were included in this study. Changes in cardiac output were provoked by body position change at rest (left lateral supine, horizontal supine, sitting, and standing) and a steady state bicycle exercise at 20 watts. Estimated CO derived from the IHM (CO(IHM)) was compared to CO measured with inert gas rebreathing (CO(RB)), echocardiography (CO(ECHO)) and impedance cardiography (CO(ICG)). CO(RB) was considered the reference method.
Results:
The median intra-patient correlation coefficient comparing CO(RB) and CO(IHM) was 0.83 (range: 0.63-0.98). Comparing CO(RB) with CO(ECHO) and CO(ICG) resulted in mean intra-patient correlation coefficients of 0.73 (-0.29-0.94) and 0.63 (-0.29-0.96). In a statistical model where slope and intercept was considered random between patients the coefficient of determination (R2) comparing CO(RB) and CO(IHM) was 0.91. Mean bias was -0.39 L/min (11%). Limits of agreement were +/-1.56 L/min and relative error was 21%.
Conclusions:
A simple algorithm based on RV pressure wave form characteristics derived from an IHM can be used to estimate changes in CO in heart failure patients. These findings encourage further research aiming to improve and validate the algorithm.
Related Concept Videos
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart rate...
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Heart Failure VII: Nursing Interventions