Phase Contrast Magnetic Resonance Imaging in the Rat Common Carotid Artery
Published on: September 5, 2018
Contrast agent influences MRI phase-contrast flow measurements in small vessels
Kerstin M Lagerstrand1, Barbro Vikhoff-Baaz, Göran Starck
1Departments of Radiation Physics, and Medical Physics and Biomedical Engineering, Göteborg University, Sahlgrenska University Hospital, Göteborg, Sweden. kerstin.lagerstrand@vgregion.se
Abstract:
Contrast-enhanced MR angiography is often combined with phase contrast (PC) flow measurement to answer a particular clinical question. The contrast agent that is administered during contrast-enhanced MR angiography may still be present in the blood during the consecutive PC flow measurement. The aim of this work was to evaluate the influence of contrast agent on PC flow measurements in small vessels. For that purpose, both in vivo measurements and computer simulations were performed. The dependence of the PC flow quantification on the signal amplitude difference between blood and stationary background tissue for various vessel sizes was characterized. Results show that the partial-volume effect strongly affects the accuracy of the PC flow quantification when the imaged vessel is small compared to the spatial resolution. A higher blood-to-background-contrast level during imaging significantly increases the partial-volume effect and thereby reduces the accuracy of the flow quantification. On the other hand, a higher blood-to-background-contrast level facilitated the segmentation of the vessel for flow rate determination. PC flow measurements should therefore be performed after contrast agent administration in large vessels, but before contrast agent administration in small vessels.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies VII: Vascular Imaging
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT