Accuracy in quantification of coronary calcification with CT: a cork-dog heart phantom study
Philip H Tseng1, Songshou Mao, David Z Chow
1Division of Cardiology, Los Angeles Biomedical Research Institute at Harbor-UCLA, Torrance, CA 90502, USA.
Rationale And Objectives:
Coronary artery calcium is a sensitive risk predictor of cardiac events. However, measurement of calcium foci is affected by partial-volume effects, which ultimately have an effect on accuracy and reproducibility of calcium scores. In this study, we describe the accuracy of quantification of calcium foci of known size and density using cork-dog heart phantoms.
Materials And Methods:
Five study phantoms were constructed from cork chests and dog hearts containing 135 calcium hydroxyapatite (CaHA) foci of known volume, mass, and concentration located in the coronary arteries or the myocardium. Hearts were separated into two groups: (1) three hearts containing large, high-density foci and (2) two hearts containing small, low-density foci. The phantoms were scanned using a standard coronary artery calcium (CAC) protocol and the volume and mean intensity of foci were measured.
Results:
In group 1, the total volume of 87 CaHA foci measured was 4284 and 3779 mm(3) with electron beam computed tomography (EBCT); multidetector computed tomography (MDCT), respectively (P < .001). Both were significantly larger than the true volume (2713.9 mm(3), P < .001). In Group 2, the total volume of 57 CaHA foci measured was 592.6 and 702.9 mm(3) with EBT and MDCT, respectively (P < .001). Both were significantly smaller than the true volume (1733.2 mm(3), P < .001). We found that EBCT values for volume were approximately generally higher than MDCT values, but strongly correlated (r = 0.95, P < .0001). Agatston scores were found to be nearly equivalent between EBCT and MDCT and were similarly strongly correlated (r = 0.97, P < .0001).
Conclusions:
Computed tomography images overestimate the volume of large, dense CaHA foci while underestimating the volume of smaller (<6.6 mm(3)), less dense foci. This may have significant implications on CAC scoring and volume measurement. EBCT overestimated calcium more than MDCT, most likely from increased image noise.
More Related Videos
08:02Novel Quantification Protocol for Cardiovascular Calcification Progression Using Longitudinal MicroPET/MicroCT Images
Published on: November 15, 2024
06:57Semi-Automatic Graphical Tool for Measuring Coronary Artery Spatially Weighted Calcium Score from Gated Cardiac Computed Tomography Images
Published on: September 22, 2023
Related Concept Videos
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Acute Coronary Syndrome III: Diagnostic Studies