Are toe pressures measured by a portable photophlethysmograph equivalent to standard laboratory tests?
Phyllis A Bonham1, Teresa Kelechi, Martina Mueller
1Wound Care Education Program, College of Nursing, Medical University of South Carolina, and Department of Vascular Surgery, Medical University Hospital, Charleston, SC 29425, USA. bonhamp@musc.edu
Purpose:
The purpose of this study was to determine if toe pressures (TPs) obtained by a registered nurse using a portable photophlethysmograph (PPG) were equivalent to TPs obtained by a registered vascular technologist (RVT) using standard laboratory equipment.
Design:
A within-subjects, comparative design was used for the study.
Setting And Subjects:
Thirty subjects referred to a vascular laboratory for arterial studies were recruited. All tests were performed in the outpatient vascular laboratory of a large, urban medical center.
Methodology:
Toe pressures were measured on subjects by the same RN and RVT during the same visit. Data were analyzed using the Bland-Altman method that compares bias (mean difference) and precision (limits of agreement) of measurements to previously determined criterion for clinically important limits of difference (±15 mm Hg) in order to assess equivalence and repeatability of measurements. Kappa statistic was calculated to assess RVT/RN agreement to detect lower extremity arterial disease (LEAD) (ie, TP<50 mm Hg). Sensitivity and specificity of the portable PPG measures were determined.
Results:
Precision for RVT-RN TPs exceeded the previously determined cutoff criteria (±15 mm Hg), but precision for repeated RN PPG measures fell within the clinically important limits. Kappa statistic calculation (κ=0.76) revealed substantial agreement (90%) between the RVT and RN measures to detect LEAD (<50 mm Hg). The portable PPG technique had good sensitivity (79%) and high specificity (95%) for detection of LEAD.
Conclusion:
Although TPs obtained by the portable PPG were not equivalent to standard laboratory tests, the portable technique agreed sufficiently with the RVT to detect LEAD. The good sensitivity and high specificity of the portable PPG make it suitable for nurses and other primary care providers to use for high-risk patients or patients with wounds, when the ankle brachial index either is elevated above 1.3 or cannot be performed. Photophlethysmograph is also suitable to assess healing potential and the need for referrals to the vascular laboratory, surgeon, or the need for adjunctive therapies.
Related Concept Videos
Measurement of Blood Pressure
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Pulse Assessment Sites
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessing Blood pressure in the Leg
Preparation:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.