Helium microwave-induced plasma mass spectrometric detection for reversed-phase high-performance liquid
D Heitkemper1, J Creed, J A Caruso
1Department of Chemistry, University of Cincinnati, OH 45221.
Abstract:
Reversed-phase high-performance liquid chromatography (HPLC) is directly coupled to helium microwave-induced plasma mass spectrometry (He MIP-MS) for the element-selective detection of halogenated organic compounds. Absolute detection limits are approximately 50 pg Br for brominated compounds, 1 pg I for iodinated compounds, and 10 ng Cl for chlorinated compounds. The linear dynamic range for Br- and I-containing compounds is 3-4 orders of magnitude. However, the linear range for chlorinated species is severely limited by high background at m/z = 35. The relative standard deviation for repetitive injections is less than 10%. The helium microwave-induced plasma is operated at moderate powers (300-350 W) and with a total helium consumption of 6-8 L/min. The effect of organic solvents on the background mass spectrum is investigated.
More Related Videos
Related Concept Videos
High-Resolution Mass Spectrometry (HRMS)
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Elution Process
High-Performance Liquid Chromatography: Instrumentation
High-Performance Liquid Chromatography: Types of Detectors