Comparative testing by cyclic loading of rotator cuff suture anchors containing multiple high-strength sutures
F Alan Barber1, Onur Hapa, James A Bynum
1Plano Orthopedic Sports Medicine and Spine Center, Plano, Texas 75093, USA.
Summary
Medial-row anchors demonstrate superior fixation strength and less displacement compared to lateral-row anchors in cadaveric models. Lateral-row anchors are more prone to failure at lower loads, highlighting differences in performance.
Area of Science:
- Orthopedic surgery
- Biomechanical testing
- Shoulder arthroscopy
Background:
- Rotator cuff tears are common, often requiring surgical repair with anchors.
- Medial-row and lateral-row anchor placement are surgical techniques used for rotator cuff repair.
- Understanding anchor performance is crucial for optimizing surgical outcomes.
Purpose of the Study:
- To compare the biomechanical performance of isolated medial-row versus isolated lateral-row anchors.
- To evaluate anchor cyclic loading and destructive testing in a cadaveric model.
- To determine failure modes and load capacities of different anchor types.
Main Methods:
- 16 human cadaveric humeri were used for anchor testing.
- Four medial-row and four lateral-row anchors were tested in different humeral head locations.
- Constructs underwent cyclic loading (10-60 N at 1 Hz for 500 cycles) followed by destructive testing.
Main Results:
- Lateral-row anchors exhibited greater displacement (4.4 mm) than medial-row anchors (1.9 mm).
- Ultimate failure loads ranged from 163 N (Footprint) to 308 N (Versalok).
- The primary failure mode was anchor pullout, with lateral-row anchors failing at lower loads and earlier cycles.
Conclusions:
- Medial-row anchors provide superior fixation strength and security compared to lateral-row anchors.
- Lateral-row anchors demonstrate increased displacement and are more susceptible to early failure at the eyelet.
- Design differences contribute to the observed performance disparities between medial and lateral row anchors.