Tribological properties of Ti-based alloys in a simulated bone-implant interface with Ringer's solution at fretting contacts
Related Experiment Video
Updated: Jun 9, 2026

An Improved Mechanical Testing Method to Assess Bone-implant Anchorage
Published on: February 10, 2014
C R Ramos-Saenz1, P A Sundaram, N Diffoot-Carlo
1Department of Mechanical Engineering, University of Puerto Rico-Mayagüez, Call Box 9000, Mayaguez, PR 00680, USA.
Abstract:
The wear properties of oxidized and non-oxidized gamma-TiAl (a potential biomaterial) as well as Ti-6Al-4V and CP-Ti disks were studied and characterized by means of standard wear tests using a custom made bone pin arrangement. The Ti-based disks were oxidized in air at 500 and 800 degrees C for one hour. The tribological properties of the oxides formed over the disks were studied using a linear reciprocating wear testing machine under both dry and simulated biological conditions using Ringer's solution. Loss of metal oxide and coefficient of friction values were determined from wear testing. From the results, abrasion and adhesion were the primary wear mechanisms in each of the three alloy-bone pairs. Specifically, the oxide formed on gamma-TiAl possessed the highest COF and wear resistance of the three materials which were studied. Also, as expected, bone wears down faster than the Ti-based metal oxide.