Corrosion behavior of plasma electrolytically oxidized gamma titanium aluminide alloy in simulated body fluid
L Lara Rodriguez1, P A Sundaram1
1Department of Mechanical Engineering, University of Puerto Rico, Call Box 9000, Mayagüez Campus, Mayagüez, PR 00681, USA.
Abstract:
Plasma electrolytic oxidized (PEO) γTiAl alloy samples were electrochemically characterized by open circuit potential (OCP), cyclic polarization and electrochemical impedance spectroscopy (EIS) to evaluate their corrosion resistance in simulated body fluid (SBF) in order to gauge their potential for biomedical applications. Experimental results through OCP and cyclic polarization studies demonstrated the protective nature and the beneficial effect of the PEO coatings on γTiAl. The PEO surface increased corrosion resistance of these surface modified alloys. EIS data indicated the presence of an underlying compact oxide layer with surface pores represented by two domes in the Nyquist plots. Electrical equivalent circuits to describe the EIS results are proposed.
More Related Videos
Related Concept Videos
Corrosion
Corrosion of Reinforcement
However, over time and under certain conditions like carbonation, chloride ingress, and cracking this protective state can be compromised. Steel has areas with...