Related Experiment Video
Updated: Jun 8, 2026

Scalable Solution-processed Fabrication Strategy for High-performance, Flexible, Transparent Electrodes with Embedded Metal Mesh
Published on: June 23, 2017
Construction of a copper stepwedge with aluminum equivalent mean gray values
Nesrin Dündar1, Pelin Güneri, Erinç Onem
1Faculty of Dentistry, Department of Oral Diagnosis and Oral Radiology, Ege University, Bornova, Izmir, Turkey. nesrinergul@yahoo.com
Objectives:
To produce a copper (Cu) stepwedge with aluminum (Al) equivalent mean gray values (MGV).
Material And Methods:
The thicknesses of Cu steps that were equivalent to those of the Al were formulated using the X-ray attenuation properties of the materials. The Al and fabricated Cu stepwedges were radiographed, but the MGVs of the Cu stepwedge were mismatching to those of the Al. Using a mathematical function to adjust the pixel MGV of Cu stepwedge to those of the Al, new Cu stepwedges were created. In vitro iterations were performed until best approximation to Al was reached.
Results:
The MGV of the Cu stepwedges fabricated by formularization were different than those of Al (p=0.001). Iteration method led to MGV similar to those of the Al stepwedge (p=0.207).
Conclusions:
Construction of a Cu stepwedge according to the basic rules of radiophysic failed to result in a stepwedge with similar radiodensity values to those of Al stepwedge. Further studies may use the formularization method only for prototype Cu wedge production, but consecutive iterations shall be compassed to obtain the best approximation to Al MGV.
Related Concept Videos
Transmission Line Design Considerations
Wedges
Consider using a wedge to lift a heavy slab. Here, the wedge functions by converting the applied force into a much larger force directed almost perpendicular to the initial force. This...