Early type III endoleak with an Endurant endograft
Issam Abouliatim1, Djelloul Gouicem, Hicham Kobeiter
1Department of Vascular Surgery, Hopital Henri-Mondor, APHP, University Paris 12, Créteil, France. i_abouliatim@hotmail.com
Journal of Vascular Surgery
|September 17, 2010
Abstract:
We report the case of an 81-year-old man who presented with an intraoperative type III endoleak after treatment with an Endurant endograft for a 60-mm abdominal aortic aneurysm. To our knowledge, this is the first case of a type III endoleak reported with this new device. It was most likely due to a tear in the polyester graft, the cause of which remains speculative. The tear was demonstrated by postoperative angiography, which was more informative than computed tomography. The endoleak was successfully treated by relining with an aorto-uni-iliac device.