Rapid and standardized quantitation of hemolytic activity of the fourth component of human complement
T Kaufmann1, H J Kytzia, C Rittner
1Institut für Rechstmedizin, Johannes-Gutenberg-Universität Mainz, FRG.
Abstract:
Based on a method that uses the fourth component of complement (C4)-deficient guinea pig serum to quantify the hemolytic activity of C4, we developed an assay that allows the processing of a large number of individual samples with high reproducibility. In contrast to the conventional procedure using titration curves of each sample to be determined, we can show that a single appropriate dilution of the sample allows accurate quantitation of hemolytic activity. The reliability of the procedure is demonstrated using either C4A- or C4B- deficient and normal individual samples.