Overcorrection of iodinated contrast attenuation in SPECT-CT: phantom studies
Dacian V Bonta1, Richard L Wahl
1Department of Nuclear Medicine, VAMC, Atlanta, Georgia, USA. dvbonta@yahoo.com
Purpose:
The authors investigated the effects of iodinated contrast agents on the quantification of radioactivity obtained using a combined single photon emission tomography (SPECT) and high resolution 16 slice computed tomography (CT) scanner.
Methods:
Two conditions were evaluated: Contrast media mixed with radioisotope and the contrast media, and the radioisotope solution confined in adjacent but distinct volumes. Phantoms containing combinations of 99mTc, 111In, and 131I solutions in normal saline and contrast solutions in normal saline corresponding to the two conditions were prepared and scanned. Images were reconstructed by iterative reconstruction (ordered subset expectation maximization), with and without CT-derived attenuation maps. The reconstructed counts encoded in the reconstructed images were compared.
Results:
Compared to normal saline, the presence of an iodinated contrast agent resulted in the underestimation of reconstructed counts in all nonattenuation corrected SPECT images. Compared to normal saline, the presence of a contrast agent resulted in the overestimation of reconstructed counts in all attenuation corrected SPECT images. The largest underestimation of reconstructed counts was 12.8% and the largest overestimation value was 35.9%.
Conclusions:
High concentrations of a contrast agent caused errors in the measurement of actual radiotracer reconstructed counts in phantoms. These errors could have important consequences for the use of oral and intravenous contrast in both diagnostic (qualitative) and dosimetric (quantitative) SPECT-CT imaging. The authors' findings suggest that noncontrast imaging or alternative contrast-specific attenuation correction approaches should be considered for optimal SPECT activity quantification.
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...