Quantitative and Visual Benefits of Data-Driven Motion Correction on Oncologic PET/CT: A Prospective Cross-sectional
Semra Ince1, M Allan Thomas1, Malak Itani1
1Department of Radiology, Washington University School of Medicine, 510 S. Kingshighway Blvd, St. Louis, MO 63110 (S.I., M.A.T., M.I., C.S., R.L., A.H., R.L.W., T.J.F.).
Elastic motion correction with deblurring (EMCD) improves oncologic PET imaging by enhancing lesion quantitation and image quality. This advanced technique offers superior results compared to conventional gating methods, benefiting FDG-PET and DOTATATE-PET scans.
Area of Science:
- Nuclear Medicine
- Medical Imaging
- Radiochemistry
Background:
- Conventional positron emission tomography (PET) respiratory gating uses a fraction of PET counts, potentially limiting image quality.
- Elastic motion correction with deblurring (EMCD) utilizes all PET counts for motion correction without increasing image noise.
Purpose of the Study:
- To evaluate the quantitative and visual effects of EMCD on FDG-PET and DOTATATE-PET images.
- To compare EMCD-corrected images against conventional optimal gate (OG) and ungated (UG) images.
Main Methods:
- Prospective study of 78 adults undergoing oncologic PET/CT with FDG or DOTATATE.
- Reconstruction of ungated (UG), belt-gating-derived optimal gate (BG-OG), belt-gating-EMCD (BG-EMCD), and data-driven gating-EMCD (DDG-EMCD) images.
- Quantitative analysis of lesion SUV-max and contrast-to-noise ratio (CNR), and visual assessment by two blinded readers.
Main Results:
- Motion-corrected images (BG-OG, BG-EMCD, DDG-EMCD) showed significantly higher lesion SUV-max than UG images (P<0.001).
- EMCD-based images (BG-EMCD, DDG-EMCD) demonstrated significantly higher CNRs compared to BG-OG and UG images (P<0.001).
- Readers preferred EMCD-based images over BG-OG images, with EMCD occasionally identifying more lesions.
Conclusions:
- EMCD-based motion correction significantly improves lesion quantitation, including higher SUVs and CNRs, and enhances overall image quality in oncologic PET.
- EMCD is a valuable tool for improving the quality of oncologic PET imaging on capable scanners.