Extended fast-ion D-alpha diagnostic on DIII-D.
C M Muscatello1, W W Heidbrink, D Taussig
1University of California, Irvine, California 92697, USA.
The Review of Scientific Instruments
|November 2, 2010
Summary
The upgraded Fast-Ion Deuterium-Alpha (FIDA) diagnostic enhances spatial coverage and velocity space measurements for energetic ions. This improved system provides crucial data for fusion energy research.
Area of Science:
- Plasma Physics
- Fusion Energy Research
- Diagnostic Techniques
Background:
- The Fast-Ion Deuterium-Alpha (FIDA) diagnostic measures energetic ions in fusion plasmas.
- The initial FIDA system was installed on DIII-D in 2005.
- Energetic ions are crucial for understanding fusion plasma dynamics.
Purpose of the Study:
- To detail the upgrades and capabilities of the second-generation (2G) FIDA diagnostic on DIII-D.
- To highlight improvements in spatial coverage and fast-ion velocity space resolution.
- To present the enhanced measurement capabilities for fusion research.
Main Methods:
- Installation of the 2G FIDA system in 2009 with 11 active in-beam and 3 passive off-beam views.
- Utilizing fast f/1.8 optical components and 1.5 mm core fibers for enhanced light gathering.
- Employing spectrometers, CCD cameras, photomultiplier tubes, and fast digitizers for data acquisition.
Main Results:
- Expanded spatial coverage with 11 active and 3 passive views.
- Enhanced fast-ion velocity space coverage due to more tangential views.
- High-frequency (1 kHz and 1 MHz) data acquisition capabilities for transient phenomena.
Conclusions:
- The 2G FIDA system significantly improves the measurement of energetic ions in fusion plasmas.
- Enhanced diagnostics are vital for advancing fusion energy science.
- The upgraded FIDA diagnostic provides more comprehensive data for plasma physics studies.
Related Concept Videos
Chemical Ionization (CI) Mass Spectrometry
The molecular ion peak of a molecule in the mass spectrum provides vital information for molecular identification. However, conventional electron impact ionization can lead to the rapid dissociation of some molecular ions before they reach the detector. A milder ionization method is required to increase the lifetime of such ionized analyte molecules. Chemical ionization (CI) is a gas-phase protonation reaction useful for mass-analyzing analyte molecules that are easily protonated to yield the...
¹³C NMR: Distortionless Enhancement by Polarization Transfer (DEPT)
When proton-coupled carbon-13 spectra are simplified by a broadband proton decoupling technique, structural information about the coupled protons is lost. Distortionless enhancement by polarization transfer (DEPT) is a technique that provides information on the number of hydrogens attached to each carbon in a molecule. While the DEPT experiment utilizes complex pulse sequences, the pulse delay and flip angle are specifically manipulated. The resulting signals have different phases depending on...