X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Comparison of the radiation dose in a cardiac IVR X-ray system
Youhei Inaba1, Koichi Chida, Kazutoshi Shirotori
1Department of Radiology, Tohoku University Hospital, 1-1 Seiryo, Aoba, Sendai 980-8574, Japan.
Abstract:
In this study, the entrance surface dose rates received by a phantom during cineangiography and fluoroscopy were compared. The X-ray conditions used in the measurements were those normally used in facilities performing percutaneous coronary intervention. Although, today, the entrance surface doses (cineangiography and fluoroscopy) of X-ray equipment used for cardiac interventional radiology (IVR) tends to be lower than they were previously, some equipment produces a high radiation dose. Therefore, the X-ray equipment used for cardiac IVR procedures must be maintained in good repair and must be carefully calibrated. In addition, periodic measurement of the radiation dose from the X-ray equipment used for both cineangiography and fluoroscopy for cardiac IVR is necessary. If the radiation dose of the X-ray system in use is too high, the IVR staff should determine the reason and make an effort to reduce it. Hence, the IVR staff must be adequately trained in radiation protection.
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System V: CT
X-ray Imaging
Imaging Studies III: Computed Tomography
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System IV: CMRI