Selected Reaction Monitoring Mass Spectrometry for Absolute Protein Quantification
Published on: August 17, 2015
Stable-labeled analogues and reliable quantification of nonprotein biomarkers by LC-MS/MS
Robert MacNeill1, Timothy Sangster, Marc Moussallie
1Department of Bioanalysis, Huntingdon Life Sciences, Woolley Road, Alconbury, Huntingdon, Cambridgeshire, UK. macneilr@ukorg.huntingdon.com
Background:
The aim was to develop, and establish as suitable to begin assessment by full validation, a quantitative LC-MS/MS method for asparagine in human plasma. Therein, to utilize a stable-labeled analogue of asparagine to act as surrogate analyte, producing complete calibration curves and corresponding QC samples and another m/z distinct stable-labeled analogue to act as internal standard.
Results:
From two candidates, the surrogate analyte was selected through statistical comparisons of concentration-response data and the resultant method employed protein precipitation and LC on an unmodified silica column with multiple reaction monitoring detection mode. The calibration range was 50-10,000 ng/ml.
Conclusion:
This method was successfully proven to meet the accuracy and precision acceptance criteria of current bioanalytical method validation guidelines.