Accuracy of single-plane fluoroscopy in determining relative position and orientation of total knee replacement
Stacey Acker1, Rebecca Li, Heather Murray
1Department of Mechanical and Materials Engineering, Queen's University, Kingston, Ontario K7L3N6, Canada. acker@me.queensu.ca
Abstract:
The accuracy of estimating the relative pose between knee replacement components, in terms of clinical motion, is important in the study of knee joint kinematics. The objective of this study was to determine the accuracy of the single-plane fluoroscopy method in calculating the relative pose between the femoral component and the tibial component, along knee motion axes, while the components were in motion relative to one another. The kinematics of total knee replacement components were determined in vitro using two simultaneous methods: single-plane fluoroscopic shape matching and an optoelectronic motion tracking system. The largest mean differences in relative pose between the two methods for any testing condition were 2.1°, 0.3°, and 1.1° in extension, abduction, and internal rotation respectively, and 1.3, 0.9, and 1.9 mm in anterior, distal, and lateral translations, respectively. For the optimized position of the components during dynamic trials, the limits of agreement, between which 95% of differences can be expected to fall, were -2.9 to 4.5° in flexion, -0.9 to 1.5° in abduction, -2.4 to 2.1° in external rotation, -2.0 to 3.9 mm in anterior-posterior translation, -2.2 to 0.4mm in distal-proximal translation and -7.2 to 8.6mm in medial-lateral translation. These mean accuracy values and limits of agreement can be used to determine whether the shape-matching approach using single-plane fluoroscopic images is sufficiently accurate for an intended motion tracking application.