MR-compatible RF ablation system for online treatment monitoring using MR thermometry
Karl Will1, Johannes Krug, Kerstin Jungnickel
1Department of Signal Processing and Communications, Otto-von-Guericke-University of Magdeburg, Universitaetsplatz 2, 39106, Germany. karl.will@oygu.de
Abstract:
RF ablation (RFA) is used for thermal ablation of tumors in which the RF electrode is placed in the tissue under image-guidance. Because of the good tumor visibility and the lack of ionizing radiation, MR-guided RFA is the method of choice. Additionally, with the help of MR thermometry the RF ablation can be monitored during the intervention. Unfortunately, the imaging of an MR scanner is highly sensitive to interferences caused by external electrical signals. In this paper the high-power RF ablation signal of a commercially available medical therapy device is made MR-compatible. A design of a low-pass filter with high-power compatibility is presented. The filter performance is demonstrated by means of simulations and measurements.
Related Concept Videos
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...