Related Experiment Video
Updated: May 1, 2026

Cell-free Biochemical Fluorometric Enzymatic Assay for High-throughput Measurement of Lipid Peroxidation in High Density Lipoprotein
Published on: October 12, 2017
[Determination of homocysteine level in plasma by high performance liquids chromatography with fluorescence
1Institute of Health and Environmental Medicine, Academy of Military Medical Sciences, Tianjin 300050, China.
Aim:
To establish a method for detection of plasma total homocysteine with HPLC.
Methods:
The chromatography analysis was carried out using a Symmetry Shield RP18. The mobile phase was sodium acetate (0.08 mol/L) and methanol (1%) and we utilized a HPLC system with fluorescence detection of plasma homocysteine derivatized from reaction with 7-fluorobenzo-2-oxa-1,3-diazole-4-sulphonate (SBD-F).
Results:
The average recoveries were 95.8 - 100.8% and the relative standard deviations were 1.2-2.0%.
Conclusion:
The results showed it to be a rapid and accurate method for the determination of homocysteine level in plasma.
More Related Videos
Related Concept Videos
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Instrumentation
High-Performance Liquid Chromatography: Types of Detectors