[Effects of 50 Hz-filter circuit on cardiac action potential recording and analyzing]
Xiao-lin Yi1, Zhen-zhong Fan, Rui-rong He
1Department of Physics, Hebei Medical University, Shijiazhuang 050017, China.
Aim:
To study the effects of the 50 Hz-filter circuit in a microelectrode amplifier on cardiac action potential waveform and parameters.
Methods:
Cardiac action potential signals were fed into a microcomputer through a glass microelectrode, a microelectrode amplifier, a differentiator and A/D converter. The cardiac action potential signals were recorded and analyzed with 50 Hz-filter circuit and without it, and the frequency spectrum in the signals was analyzed with the fast Fourier transformation.
Results:
When the 50 Hz-filter circuit was used, the phase 0 of the potential waveform was seriously distorted and prolonged. The maximal rate of depolarization at the phase 0 was cut down, while the other parameters were not effected.
Conclusion:
There has already been much 50 Hz element in the action potential waveform. During amplifying the cardiac action potential signal, the 50 Hz-filter circuit should not be turned on. Otherwise, the experiment results will be effected.