Related Experiment Video
Updated: Jun 5, 2026

Using Extraordinary Optical Transmission to Quantify Cardiac Biomarkers in Human Serum
Published on: December 13, 2017
Development of a surface plasmon resonance-based immunosensor for the rapid detection of cardiac troponin I
Young-Chul Kwon1, Min-Gon Kim, Eon-Mi Kim
1Laboratory of Microbiology of GNU Medical School, Gyeongsang National University, 900 Gajwa-Dong, Chinju, Korea.
Abstract:
The concentration of cardiac troponin I (cTnI) in blood is an important marker for heart muscle cell damage. A surface plasmon resonance (SPR)-based immunosensor was devised for the rapid and specific detection of cTnI. It was constructed by crosslinking a monoclonal antibody P-II-13, which was generated against a loop region (aa 84-94) of cTnI protein as an epitope peptide, onto a chemically modified thin gold film. The performance of the sensor was examined with respect to the SPR signal intensity versus cTnI concentration. The signal intensity was directly correlated with the cTnI concentration in the range of 0-160 μg/l. The sensor signal was saturated when the concentration of cTnI approached 660 μg/l with the SPR intensity of 172 RU. The lower detection limit of the sensor was 68 ng/l cTnI, which was comparable to ELISA-based commercial cTnI detection systems.
More Related Videos
08:58Fully Automated Centrifugal Microfluidic Device for Ultrasensitive Protein Detection from Whole Blood
Published on: April 16, 2016
10:12A Sensitive and Specific Quantitation Method for Determination of Serum Cardiac Myosin Binding Protein-C by Electrochemiluminescence Immunoassay
Published on: August 8, 2013