[Clinical application of prospective electrocardiogram-gated 320-detector computed tomography coronary angiography]
Jie Qin1, Ling-Yun Liu, Xiao-Chun Meng
1Department of Radiology, Department of Cardiology, the Third Affiliated Hospital of Sun Yat-Sen University, Guangzhou 510630, China.
Objective:
to compare the success rate, radiation dose, image quality and diagnosis of prospective electrocardiogram(ECG)-gated 320-detector computed tomography coronary angiography (CTCA) versus retrospective ECG-gated CTCA.
Methods:
patients suspected coronary artery disease were divided into two groups which underwent 320-detector CTCA with prospective ECG-gated and retrospective ECG-gated scanning (n = 240 each, HR < 65 bpm). Curved-planar reconstruction (CPR), maximum intensity projection (MIP) and volume rendering (VR) were performed to demonstrate the coronary arteries. The image quality was defined as excellent, good and poor by motion and stair-step artifacts. Effective radiation exposure dose was estimated from the dose-length product. Effective radiation dose, image quality and diagnosis were evaluated.
Results:
the success rate of examination was 100% in prospective ECG-gated group and retrospective ECG-gated group. The mean effective radiation dose of prospective ECG-gated CTCA [(3.3 ± 1.3) mSv] was significantly lower than that of retrospective ECG-gated CTCA [(13.0 ± 1.6) mSv, P < 0.01]. Segments of diagnostic image quality (95.42%, 3435/3600) and non-diagnostic coronary segments (4.58%, 165/3600) in prospective ECG-gated group were similar as those of retrospective ECG-gated group (95.81%, 3449/3600 and 4.19%, 151/3600, all P > 0.05). Compared with CAG, the sensitivity, specificity, false positive and false negative value in prospective ECG-gated group (93.22%, 99.21%, 91.64%, 99.05%) and retrospective ECG-gated group (94.55%, 98.80%, 95.86%, 98.54%) were not significantly different.
Conclusion:
though the effective radiation dose is significantly lower, the success rate, image quality and diagnosis of prospective ECG-gated 320-detector CTCA is comparable with that of retrospective ECG-gated 320-detector CTCA on patients with stable heart rates less than 65 bpm.
More Related Videos
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Acute Coronary Syndrome III: Diagnostic Studies
Cardiac Catheterization I: Pre-Procedure Overview
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...