Closed-loop control of mechanical ventilation: description and classification of targeting schemes
Robert L Chatburn1, Eduardo Mireles-Cabodevila
1Respiratory Institute, The Cleveland Clinic, and with Lerner College of Medicine, Case Western Reserve University, Cleveland, Ohio 44195, USA. chatbur@ccf.org
Abstract:
There has been a dramatic increase in the number and complexity of new ventilation modes over the last 30 years. The impetus for this has been the desire to improve the safety, efficiency, and synchrony of ventilator-patient interaction. Unfortunately, the proliferation of names for ventilation modes has made understanding mode capabilities problematic. New modes are generally based on increasingly sophisticated closed-loop control systems or targeting schemes. We describe the 6 basic targeting schemes used in commercially available ventilators today: set-point, dual, servo, adaptive, optimal, and intelligent. These control systems are designed to serve the 3 primary goals of mechanical ventilation: safety, comfort, and liberation. The basic operations of these schemes may be understood by clinicians without any engineering background, and they provide the basis for understanding the wide variety of ventilation modes and their relative advantages for improving patient-ventilator synchrony. Conversely, their descriptions may provide engineers with a means to better communicate to end users.
Related Concept Videos
Mechanical Ventilation I: Indication and Settings
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Open and closed-loop control systems
An open-loop control system operates without feedback from the output. It consists of two primary elements: the controller and the controlled process. The controller receives an input signal and...
Cardiopulmonary Resuscitation II: ACLS Airway Management