Comparison of two ventilator circuits for Dräger Babylog high-frequency ventilation
Kevin I Wheeler1, Gregory P Moore, Colin J Morley
1Newborn Services, Royal Women's Hospital Monash, Melbourne, Victoria, Australia. kevin.wheeler@thewomens.org.au
Aim:
The Dräger Babylog 8000plus ventilator (Dräger Medical Systems, Lübeck, Germany) can provide both conventional and high-frequency ventilation (HFV). Dräger recommends specific circuits for each of these modes. We investigated the performance of the Babylog ventilator in HFV mode when used with the recommended circuits for both conventional and HFV.
Methods:
The Fisher and Paykel RT235 (conventional; Fisher and Paykel Healthcare, Auckland, New Zealand) and Hytrel (HFV; Fisher and Paykel Healthcare) circuits were studied using a 50-mL test lung. Tidal volume, high-frequency minute volume and ventilator alarms were compared at 100 combinations of mean airway pressures (10-16 cm H₂O), frequencies (6-14 Hz) and amplitudes (20-60%).
Results:
Tidal volume with the two circuits differed by < 5% for tidal volumes ≤ 2.5 mL. Above this, tidal volumes delivered with the HFV circuit were up to 15% more than that obtained with the conventional ventilation circuit, and high-frequency minute volume differed by up to 30%. With the exception of the highest tidal/minute volumes, the tidal volume delivered using the HFV circuit could also be achieved with adjusted frequency or amplitude when using the conventional circuit. More 'pressure measurement out of range' alarms were noted with the conventional ventilation circuit, particularly at mean airway pressure ≥ 14 cm H₂O and frequency ≤ 10 Hz.
Conclusions:
The conventional ventilation circuit may allow delivery of adequate tidal volume for some infants. Where requirements are higher, the HFV circuit allows the Babylog to deliver higher tidal volumes and higher minute volume, and reduce alarms.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation I: Indication and Settings
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...