Related Experiment Video
Updated: Jun 5, 2026

Magnetic Resonance Imaging of Multiple Sclerosis at 7.0 Tesla
Published on: February 19, 2021
Reducing SAR and enhancing cerebral signal-to-noise ratio with high permittivity padding at 3 T
Qing X Yang1, Jianli Wang, Jinghua Wang
1Departments of Radiology, The Pennsylvania State University College of Medicine, 500 University Drive, Hershey, Pennsylvania, USA. qyang@psu.edu
Abstract:
Previous works have shown that placement of a high-dielectric pad can improve image intensity in a region adjacent to the pad, or that placement of dielectric pads around a large surface of the head can improve image homogeneity on an entire plane through the head in high-field MRI. Here, experimental results show that use of high-dielectric pads around the human head can reduce the required input radiofrequency power by 50% while enhancing image signal-to-noise ratio by 20-40% throughout the cerebrum at 3 T. Thus, dielectric pads may be used to provide a relatively simple and low-cost method for improving quality and safety of MRI in a variety of applications at 3 T.