[Metrological controls for impression tonometers - ensuring quality standards in ophthalmology]
1Physikalisch-Technische Bundesanstalt, AG 8.13, Berlin. thomas.schwenteck@ptb.de
Summary
Metrological controls ensure impression tonometer accuracy for eye care. Test devices are generally suitable, but regular recalibration is crucial for reliable intraocular pressure measurements.
Area of Science:
- Ophthalmology
- Metrology
- Medical Device Engineering
Background:
- Impression tonometers measure intraocular pressure (IOP) for glaucoma diagnosis and surgical monitoring.
- Mandatory metrological controls in Germany ensure tonometer accuracy and patient safety.
- Specialized test devices developed by Physikalisch-Technische Bundesanstalt (PTB) are used for these controls.
Purpose of the Study:
- To evaluate the suitability and accuracy of existing test devices for impression tonometers.
- To assess the performance of these devices against established metrological standards.
- To ensure the reliability of IOP measurements in clinical and surgical settings.
Main Methods:
- Investigated 18 test devices from various German metrology and testing institutions.
- Assessed measurement accuracy and suitability for metrological control procedures.
- Analyzed parameters including tonometer mass, plunger-lever-pointer system mass, and plunger displacement using statistical methods.
Main Results:
- The majority of investigated test devices are suitable for metrological controls of impression tonometers.
- Most devices performed within permissible error tolerances for IOP measurement.
- The reference tonometer itself showed a deviation in one parameter, which was accurately identified by all test devices.
Conclusions:
- Test devices for impression tonometers are generally reliable when properly maintained.
- Regular recalibration, at least every three years, is essential to maintain device accuracy.
- Even older devices (20+ years) remain fit for purpose with appropriate maintenance and calibration.
Related Concept Videos
Quality Control
Quality control is one of the three cyclical quality assurance activities that help keep a system under statistical control. Typical quality control activities include creating quality control charts, conducting proficiency testing, and documenting and archiving results.
Quality control helps track data, visualize trends, and identify variations, making it easier to detect deviations that may affect the accuracy of an analysis. One way to do this is by generating a quality control chart, which...
Quality control helps track data, visualize trends, and identify variations, making it easier to detect deviations that may affect the accuracy of an analysis. One way to do this is by generating a quality control chart, which...
Quality Assurance
Quality assurance is the overarching term used to describe the activities employed to ensure the proper performance of a system. These activities can be classified into three categories: quality control, quality assessment, and internal corrective measures. Typically, these activities work cyclically: quality control is performed before and during the analysis, while quality assessment occurs during and after the investigation. Internal corrective measures are implemented based on the findings...
Uncertainty in Measurement: Reading Instruments
Counting is the type of measurement that is free from uncertainty, provided the number of objects being counted does not change during the process. Such measurements result in exact numbers. By counting the eggs in a carton, for instance, one can determine exactly how many eggs are there in the carton. Similarly, the numbers of defined quantities are also exact. For example, 1 foot is exactly 12 inches, 1 inch is exactly 2.54 centimeters, and 1 gram is exactly 0.001 kilograms. Quantities...
Instrument Calibration
Instrument calibration is essential for ensuring that instruments produce accurate and consistent results. It is vital in manufacturing, healthcare, testing laboratories, and scientific research. Calibration processes are specific to each instrument and help enhance data accuracy. Each instrument has a unique calibration process tailored to its design and function to improve data accuracy.
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Glassware Calibration
Accurate calibration of glassware, such as volumetric flasks, pipettes, and burettes, is essential to ensure accurate measurements in the analytical laboratory. Calibration helps maintain consistency across measurements and prevents errors arising from inaccurate volumes.
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Bioreactor Controls-I
Maintaining optimal conditions within fermenters is essential for maximizing microbial productivity and ensuring process efficiency. This lesson focuses on key parameters—temperature, foam, pH, carbon dioxide, oxygen, and pressure—and their precise measurement and control strategies in fermentation systems.Temperature ControlTemperature regulation is critical due to the exothermic nature of many fermentation processes. In small laboratory fermenters, temperature is commonly monitored using...