Calibration-free device sizing using an inverse geometry x-ray system.

Michael T Tomkowiak1, Michael A Speidel, Amish N Raval

  • 1Department of Biomedical Engineering, University of Wisconsin-Madison, Madison, Wisconsin 53705, USA.

Medical Physics
|March 3, 2011
PubMed
Summary

A new scanning-beam digital x-ray (SBDX) system accurately measures vessel dimensions for cardiovascular interventions. This tomosynthetic imaging method eliminates the need for calibration and works even with foreshortened vessels.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Safety and Efficacy of Histotripsy Treatments Near Implanted Metallic Fiducials and Endovascular Coils in Phantoms and in an In Vivo Swine Model.

Ultrasound in medicine & biology·2026
Same author

Injectable radiopaque targets for cone-beam CT guided histotripsy.

International journal of hyperthermia : the official journal of European Society for Hyperthermic Oncology, North American Hyperthermia Group·2026
Same author

A simple clinical scoring model to predict significant left main/three-vessel disease in non-ST elevation acute coronary syndrome: The MADISON score.

International journal of cardiology·2025
Same author

Accuracy and reproducibility of a single-pose image-to-robot registration method for mobile C-arm cone beam CT guided histotripsy.

Journal of applied clinical medical physics·2025
Same author

Technique selection and technical developments for 2D dual-energy subtraction angiography on an interventional C-arm.

Medical physics·2025
Same author

Intermediate monocytes in blood correlate with subclinical vascular changes in lupus nephritis.

Lupus science & medicine·2025

Area of Science:

  • Medical Imaging
  • Interventional Cardiology
  • Biomedical Engineering

Background:

  • Quantitative coronary angiography (QCA) is crucial for device selection in cardiovascular interventions.
  • Conventional QCA accuracy relies on reference object calibration and avoiding vessel foreshortening.
  • Limitations in QCA hinder precise measurements in complex anatomies.

Purpose of the Study:

  • To introduce and validate a novel device sizing method using inverse geometry scanning-beam digital x-ray (SBDX) fluoroscopy.
  • To enable accurate measurement of vessel diameter and length without reference calibration or foreshortening concerns.
  • To overcome limitations of conventional QCA in interventional procedures.

Main Methods:

  • SBDX system generates tomosynthetic x-ray images, enabling 3D localization of vessel edges via blurring analysis.

More Related Videos

Tree Core Analysis with X-ray Computed Tomography
06:56

Tree Core Analysis with X-ray Computed Tomography

Published on: September 22, 2023

Measuring 3D In-vivo Shoulder Kinematics using Biplanar Videoradiography
06:09

Measuring 3D In-vivo Shoulder Kinematics using Biplanar Videoradiography

Published on: March 12, 2021

Related Experiment Videos

Last Updated: Jun 4, 2026

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
08:30

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging

Published on: September 11, 2011

Tree Core Analysis with X-ray Computed Tomography
06:56

Tree Core Analysis with X-ray Computed Tomography

Published on: September 22, 2023

Measuring 3D In-vivo Shoulder Kinematics using Biplanar Videoradiography
06:09

Measuring 3D In-vivo Shoulder Kinematics using Biplanar Videoradiography

Published on: March 12, 2021

  • A 3D vessel centerline is constructed to determine magnification and angulation.
  • Model-based diameter and centerline-derived length measurements were validated using phantoms with varying dimensions and orientations.
  • Main Results:

    • Phantom studies demonstrated high accuracy: average diameter errors <0.15 mm and length errors <1% (0.3 mm).
    • Measurement accuracy was maintained across various magnifications (87%-118%) and foreshortening angles (0°-75°).
    • Diameter errors were minimal (<0.25 mm) across angulations, with increased accuracy for smaller vessels and lower angles.

    Conclusions:

    • Tomosynthetic imaging with SBDX provides accurate vessel dimension measurements independent of magnification and angulation.
    • The SBDX method offers a potentially more accurate and convenient alternative to conventional QCA.
    • This technology may enhance precision and efficiency in cardiovascular interventions.