False positive troponin result caused by a true macrotroponin
Etienne C H J Michielsen1, Pieter G T Bisschops, Marcel J W Janssen
1Department of Clinical Chemistry and Hematology, VieCuri Medical Centre, Venlo, The Netherlands. etiennemichielsen@gmail.com
Background:
A 35-year-old female presented with complaints of tiredness, shortness of breath and recent onset of chest pain. The cardiac troponin I (cTnI) concentration was elevated at 6.4 μg/L (Abbott Architect reference value: <0.03 μg/L) with a normal ECG. Physical examination, radiological imaging and routine laboratory investigations did not provide an explanation for the elevated cTnI concentration.
Methods:
Use of different troponin assays, dilutions with assay diluent, addition of mouse serum, polyethylene glycol (PEG) precipitation and gel filtration chromatography (GFC) were used to investigate possible interference.
Results:
Troponin concentrations were below the level of detection when measured using all other troponin immunoassays. Serial dilutions of sample and addition of mouse serum did not alter the results. However, PEG precipitation and GFC showed the presence of a high molecular weight immunoreactive protein. Using GFC and protein-A IgG precipitation, the interference could be identified as a macrocomplex containing IgG and (fragments of) cTnI.
Conclusions:
We report a false positive cTnI result caused by a true macrotroponin, containing IgG and (fragments of) cTnI. This macrotroponin was only immunoreactive in the Abbott Architect cTnI immunoassay. Clinicians should be aware of analytical interference when troponin results are constantly elevated in the absence of (non)coronary causes of an increased troponin.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Blood Studies for Cardiovascular System I: Cardiac Biomarkers
The essential diagnostic tools for detecting myocardial necrosis and monitoring individuals suspected of having acute coronary syndrome (ACS) include:
Troponins
Troponins, particularly cardiac troponins I and T, are the most precise and sensitive markers of myocardial injury. They are detectable within 4-6 hours of myocardial injury and remain...
Acute Coronary Syndrome I: Introduction
Errors In Hypothesis Tests
Myocarditis II: Clinical Features and Diagnostic Tests
Pulmonary Tuberculosis IV
Several diagnostic approaches are used to detect TB. The conventional method is the Tuberculin Skin Test (TST), also known as the Mantoux test. However, this method has...