Related Experiment Video
Updated: Jun 3, 2026

Application of Biochip Microfluidic Technology to Detect Serum Allergen-specific Immunoglobulin E (sIgE)
Published on: April 21, 2019
A rapid gel electrophoretic chip for serum cholesterol determination
Kenta Kaminikado1, Ryuzo Ikeda, Koutaro Idegami
1School of Materials Science, Japan Advanced Institute of Science and Technology, 1-1 Asahidai, Nomi City, Ishikawa 923-1292, Japan.
This study introduces a rapid gel electrophoretic chip for simultaneous serum cholesterol determination, accurately measuring low-density lipoprotein cholesterol (LDL-C) and high-density lipoprotein cholesterol (HDL-C) levels.
Area of Science:
- Biochemistry
- Analytical Chemistry
- Biotechnology
Background:
- Serum cholesterol levels, particularly low-density lipoprotein cholesterol (LDL-C) and high-density lipoprotein cholesterol (HDL-C), are critical indicators of cardiovascular health.
- Conventional methods for cholesterol determination can be time-consuming and require specialized laboratory equipment.
- There is a need for rapid, efficient, and accurate methods for simultaneous lipoprotein analysis.
Purpose of the Study:
- To develop and validate a novel gel electrophoretic chip for the rapid determination of serum cholesterol.
- To assess the chip's capability for simultaneous quantification of LDL-C and HDL-C.
- To compare the chip-based method with a conventional enzyme-based method for cholesterol analysis.
Main Methods:
- A gel electrophoretic chip was fabricated using a photolithography technique with a 2.5% acrylamide and 1% agarose gel.
- Serum samples were diluted, stained, and loaded onto the chip for a 15-minute separation.
- Lipoprotein band intensities were analyzed using an image analyzer, and concentrations were correlated with a standard enzyme method.
Main Results:
- The electrophoretic chip demonstrated linear detection ranges for LDL-C up to 146 mg/dL and HDL-C up to 53 mg/dL.
- Analysis of real biological samples from nine volunteers showed strong correlations with the conventional enzyme-based method (r² = 0.98 for LDL-C, r² = 0.86 for HDL-C).
- The chip successfully determined lipoprotein concentrations simultaneously.
Conclusions:
- The developed gel electrophoretic chip offers a rapid and accurate method for simultaneous determination of serum LDL-C and HDL-C.
- This proof-of-concept study validates the potential of chip-based electrophoresis for clinical lipoprotein analysis.
- Further studies with larger and diverse populations are warranted to fully establish its clinical utility.
Related Concept Videos
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Two-dimensional Gel Electrophoresis
The first dimension separation uses the isoelectric focusing or IEF technique performed on immobilized pH gradient (IPG) strips that separate proteins according to their isoelectric points.
Biological samples, such as cells...
Enzyme-Linked Immunosorbent Assay
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or quantified.
Capillary Electrophoresis: Instrumentation