Horizontal pull-out strength of orthodontic infrazygomatic mini-implant: an in vitro study
Chun-Ming Chen1, Ju-Hui Wu, Pei-Chen Lu
1Faculty of Dentistry, College of Dental Medicine, Kaohsiung Medical University, University Hospital, No. 100 Shih-chuan 1st Road, Kaohsiung, Taiwan.
Purpose:
New modified mini-implants have recently come into use for reinforcing skeletal anchorage in orthodontic application. The aim of this study was to investigate the influence of the design of a mini-implant on its mechanical strength.
Materials And Methods:
We measured the insertion torques and horizontal pull-out strengths of 3 brands of infrazygomatic mini-implants (AbsoAnchor, Bioray, and Lomas; 2 mm for all). Five implants of each brand were manually driven 6 mm into the artificial bone. Significant differences in various parameters among the brands were investigated with the Kruskal-Wallis test.
Results:
There was no significant relationship between insertion torque and horizontal pull-out strength. The Bioray mini-implants had significantly greater horizontal pull-out strength than the AbsoAnchor mini-implants.
Conclusions:
The design of the mini-implant can influence its insertion torque and horizontal pull-out strength. In our findings, the horizontal pull-out strength of all mini-implants placed in the infrazygomatic crest was significantly greater than the orthodontic force applied.