Suppression of vascular enhancement artifacts through the use of a multiband, selectively spoiled radiofrequency
Dustin K Ragan1, James A Bankson
1Department of Imaging Physics, The University of Texas M.D. Anderson Cancer Center, Houston, Texas 77030-4009, USA.
Purpose:
To test the ability of a multi-band RF pulse to reduce flow enhancement artifacts for steady state imaging without compromising temporal resolution or spatial coverage.
Materials And Methods:
Selectively spoiled composite RF pulses that provide simultaneous excitation and flow preparation were designed and tested by means of simulation, phantom, and in vivo measurements under varying conditions of flow.
Results:
Suppression of flow enhancement was found to depend on flow velocity and spatial extent of spoiled regions. By determining necessary pulse characteristics for a given experimental geometry, flow enhancement was reduced and sensitivity to T(1)-reducing contrast agent was dramatically increased.
Conclusion:
These pulses provide an effective means of suppressing flow enhancement without sacrificing temporal resolution or spatial coverage.
More Related Videos
09:36Extracting Visual Evoked Potentials from EEG Data Recorded During fMRI-guided Transcranial Magnetic Stimulation
Published on: May 12, 2014
08:19Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 16, 2018
Related Concept Videos
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...