Infrared thermographic profiles of vessel sealing devices on thyroid parenchyma
Carter T Smith1, Barbara Zarebczan, Amal Alhefdhi
1Department of Surgery, University of Wisconsin School of Medicine and Public Health, Madison, Wisconsin 53792, USA.
Background:
During thyroid lobectomy, division of the thyroid parenchyma has traditionally been accomplished using suture ligation. Development of hemostatic techniques in the forms of ultrasonic dissection (UD) and electronic vessel sealing (EVS) have increased the usage of these devices during thyroid operations. We sought to characterize the thermal profile of each of these devices when used to divide the parenchyma of the thyroid gland.
Methods:
Using a porcine model, the parenchyma of the gland was sealed by alternating application of the UD and EVS devices. In each case, the thermal activity was recorded using infrared thermal imaging. We performed multiple seals with each instrument and then compared the thermal profiles.
Results:
There was no significant difference in lateral thermal spread of EVS and UD above 39, 40 or 60°C (2.30 ± 0.31 mm versus 2.53 ± 0.47 mm, P = 0.26; 2.22 ± 0.27 mm versus 2.47 ± 0.47 mm, P = 0.22, and 1.37 ± 0.27 mm versus 1.54 ± 0.26 mm, P = 0.22). There was no significant difference in mean time above 39 or 40°C (35.1 ± 8.7 s versus 31.7 ± 9.3 s, P = 0.47 and 29.9 ± 8.1 s versus 27.3 ± 6.7 s, P = 0.50). UD reached a greater maximum temperature (179.12 ± 0.0008C versus 96.52 ± 5.6C, P ≤ 0.001) and stayed over 60°C for longer than EVS (9.5 ± 1.8 s versus 5.3 ± 0.97 , P ≤ 0.001).
Conclusions:
The amount of lateral spread of thermal energy was not significantly different between the UD and EVS devices. However, the use of UD produced a higher maximum temperature during thyroid parenchyma sealing and remained above 60°C longer than EVS. This may translate into greater thermal injury to thyroid and surrounding tissues during division.
More Related Videos
11:46Investigation of the Electrophysiological and Thermographic Safety Parameters of Surgical Energy Devices During Thyroid and Parathyroid Surgery in a Porcine Model
Published on: October 13, 2022
06:08Quantitative Visualization and Detection of Skin Cancer Using Dynamic Thermal Imaging
Published on: May 5, 2011
Related Concept Videos
The Thyroid Gland
The follicles have a central cavity lined by simple cuboidal to squamous epithelial cells called follicular cells. These cells produce the glycoprotein...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Imaging Studies II: Ultrasonography