An enzymic and centrifugal method for estimating high-density lipoprotein cholesterol
Clinical Chemistry
|February 1, 1979
Summary
This study introduces a new enzymic method for measuring high-density lipoprotein cholesterol using polyethylene glycol precipitation. This approach simplifies analysis and offers accurate, cost-effective results for both men and women.
Area of Science:
- Clinical Chemistry
- Biochemistry
- Lipid Metabolism
Background:
- Accurate measurement of high-density lipoprotein cholesterol (HDL-C) is crucial for cardiovascular risk assessment.
- Traditional precipitation methods for HDL-C determination often involve complex procedures and limitations.
Purpose of the Study:
- To describe a novel enzymic method for measuring HDL-C using a centrifugal analyzer.
- To evaluate the efficacy of polyethylene glycol (PEG) for precipitating interfering lipoproteins.
Main Methods:
- Utilized polyethylene glycol (Mr 6000) at a final concentration of 100 g/L to precipitate low-density and very-low-density lipoproteins.
- Employed ethylenediaminetetraacetate-stabilized plasma, overcoming challenges associated with heparin/Mn2+ precipitation.
- Verified complete precipitation of beta-lipoproteins and retention of alpha-lipoproteins in solution using rocket immunoelectrophoresis.
Main Results:
- Achieved excellent reproducibility with a between-run coefficient of variation (CV) of 3.6% and a within-run CV of 0.8%.
- Demonstrated cost-effectiveness with reagent costs at $US 0.13 per test and high sample throughput capability.
- Established normal ranges for HDL-C: 1.20 +/- 0.31 mmol/L for men and 1.52 +/- 0.38 mmol/L for women (n=539 men, n=444 women).
Conclusions:
- The described enzymic method using PEG precipitation offers a reliable, efficient, and economical approach for HDL-C measurement.
- This method simplifies the analysis of HDL-C, making it suitable for routine clinical laboratory use.
- Provides updated normal reference ranges for HDL-C in adult men and women.