An Improved Mechanical Testing Method to Assess Bone-implant Anchorage
Published on: February 10, 2014
Characterization methods of bone-implant-interfaces of bioresorbable and titanium implants by fracture mechanical
E K Tschegg1, R A Lindtner, V Doblhoff-Dier
1Institute for Building Construction and Technology E-206-4, Vienna University of Technology, Karlsplatz 13, 1040 Vienna, Austria.
Abstract:
Bioresorbable materials for implants have become increasingly researched over the last years. The bone-implant-interfaces of three different implant materials, namely a new bioresorbable magnesium alloy, a new self-reinforced polymer implant and a conventional titanium alloy, were tested using various methods: push-out tests, SEM and EDX analyses as well as surface analyses based on stereoscopic 3D pictures were conducted. The fracture energy is proposed as a very significant reference value for characterizing the mechanical performance of a bone-implant system. By using a video-extensometer system instead of, as is commonly done, tracking the movement of the crosshead in the push-out tests, the accuracy of measurement could be increased.