Gas Chromatography-Mass Spectrometry Paired with Total Vaporization Solid-Phase Microextraction as a Forensic Tool
Published on: May 25, 2021
[Determination of tetrahydrofuran in urine by headspace solid-phase microextraction and gas chromatography]
Zhi-hui Han1, Sheng Fu, Ke-ping Yu
1Institute of Labour Hygiene and Occupational Diseases of Hunan Province, Changsha, Hunan Province 410007, China.
Objective:
Headspace solid-phase microextraction (HS-SPME) was used pre-concentration procedure for the determination of tetrahydrofuran in urine by gas chromatography with hydrogen flame detector.
Methods:
Several parameters controlling SPME was studied and optimised: SPME fiber, extraction time and extraction temperature, desorption time and desorption temperature.
Results:
Under optimal conditions, the correlation coefficient was 0.9998 and good recoveries (range from 93.0% ∼ 100.8%) were achieved, the detection limit was 0.5 µg/L.
Conclusion:
The method can be applied to the determination of trace amount of tetrahydrofuran in urine.