Related Experiment Video
Updated: Jun 1, 2026

08:08
qPCR Is a Sensitive and Rapid Method for Detection of Cytomegaloviral DNA in Formalin-fixed, Paraffin-embedded Biopsy Tissue
Published on: July 9, 2014
Precision across the analytical measuring range of a quantitative real-time PCR assay for cytomegalovirus detection
Thomas E Grys1, Doreen L Duquette, Bruce White
1Molecular Microbiology Laboratory, Mayo Medical Laboratories New England, 160 Dascomb Rd., Andover, MA 01810, USA. grys.thomas@mayo.edu
Journal of Clinical Microbiology
|June 10, 2011
Abstract:
This study measured the precision of a quantitative laboratory-developed real-time PCR test for cytomegalovirus performed at three different clinical laboratories that use the same methodology. The overall standard deviation (adjusted for analyte level) was 0.18 log(10) copies/ml, and there was no significant relationship between standard deviation and analytical measuring range.