Cardiac-induced physiological noise in 3D gradient echo brain imaging: effect of k-space sampling scheme
Anders Kristoffersen1, Pål Erik Goa
1MI Lab, Department of Medical Imaging, St. Olavs Hospital HF, N-7006 Trondheim, Norway. Anders.Kristoffersen@stolav.no
Abstract:
The physiological noise in 3D image acquisition is shown to depend strongly on the sampling scheme. Five sampling schemes are considered: Linear, Centric, Segmented, Random and Tuned. Tuned acquisition means that data acquisition at k-space positions k and -k are separated with a specific time interval. We model physiological noise as a periodic temporal oscillation with arbitrary spatial amplitude in the physical object and develop a general framework to describe how this is rendered in the reconstructed image. Reconstructed noise can be decomposed in one component that is in phase with the signal (parallel) and one that is 90° out of phase (orthogonal). Only the former has a significant influence on the magnitude of the signal. The study focuses on fMRI using 3D EPI. Each k-space plane is acquired in a single shot in a time much shorter than the period of the physiological noise. The above mentioned sampling schemes are applied in the slow k-space direction and noise propagates almost exclusively in this direction. The problem then, is effectively one-dimensional. Numerical simulations and analytical expressions are presented. 3D noise measurements and 2D measurements with high temporal resolution are conducted. The measurements are performed under breath-hold to isolate the effect of cardiac-induced pulsatile motion. We compare the time-course stability of the sampling schemes and the extent to which noise propagates from a localized source into other parts of the imaging volume. Tuned and Linear acquisitions perform better than Centric, Segmented and Random.
More Related Videos
09:55Troubleshooting and Quality Assurance in Hyperpolarized Xenon Magnetic Resonance Imaging: Tools for High-Quality Image Acquisition
Published on: January 5, 2024
11:13Quantification of Mouse Heart Left Ventricular Function, Myocardial Strain, and Hemodynamic Forces by Cardiovascular Magnetic Resonance Imaging
Published on: May 24, 2021
Related Concept Videos
Imaging Studies for Cardiovascular System IV: CMRI
Magnetic Resonance Imaging