Repeatability of SUV measurements in serial PET
J Schwartz1, J L Humm, M Gonen
1Department of Medical Physics, Memorial Sloan-Kettering Cancer Center, New York, New York 10021, USA. schwarj1@mskcc.org
Medical Physics
|July 23, 2011
Summary
This study quantifies statistical variations in standardized uptake value (SUV) measurements. SUV mean shows half the fluctuations of SUV max, with both becoming similar for small regions and short scan times.
Area of Science:
- Nuclear Medicine
- Medical Imaging Physics
- Quantitative PET Imaging
Background:
- Standardized uptake value (SUV) is crucial for monitoring cancer treatment response using FDG-PET/CT.
- Understanding the statistical uncertainty of SUV measurements is essential for accurate interpretation.
Purpose of the Study:
- To assess the reproducibility and statistical uncertainty of SUV max and SUV mean.
- To determine the minimum uncertainty threshold for physiological change detection in SUV.
Main Methods:
- Measurements were performed on a GE Discovery STE PET/CT scanner using a 68Ge phantom, an 18FDG phantom with varying target sizes, and eight cancer patients with healthy livers.
- Acquisition times ranged from 3 to 30 minutes for phantoms and included sequential 3-minute scans for patients.
- Repeatability coefficients and average percent changes in SUV max and SUV mean were calculated.
Main Results:
- For the 68Ge phantom, repeatability coefficients for SUV max and SUV mean decreased with longer acquisition times (3-30 min).
- In the 18FDG phantom, SUV max and SUV mean fluctuations decreased with increasing target diameter.
- Patient liver scans showed variability in SUV max and SUV mean repeatability, with overall ROI repeatability ranging from 27.9% to 41.1%.
Conclusions:
- SUV mean exhibits approximately half the statistical fluctuation of SUV max.
- For short scan durations (approx. 3 min) and small regions, SUV max and SUV mean have similar statistical fluctuations.
- Increased scanning time or larger target regions significantly reduce the statistical fluctuations of SUV mean compared to SUV max.