Related Experiment Video
Updated: May 30, 2026

Extracellular Glucose Depletion as an Indirect Measure of Glucose Uptake in Cells and Tissues Ex Vivo
Published on: April 6, 2022
Non-invasive in vitro sensing of D-glucose in pig blood
Harutyun Melikyan1, Emma Danielyan, Seungwan Kim
1Department of Physics and Basic Science Institute for Cell Damage Control, Sogang University, Seoul, Republic of Korea.
Abstract:
We have developed an electromagnetic resonant spiral sensor and have measured the glycemia in pig blood and the concentration of D-glucose in aqueous solution by using a real-time electromagnetic interaction phenomenon between the microwave sensor and the liquid. We could determine the concentration of glucose with a minimal resolution of 5 mg/dl in the 100-600 mg/dl concentration range at operating frequencies of about 7.65 GHz (for the glucose aqueous solution) and 7.77 GHz (for the pig blood sample). The change in the glucose concentration brings the changes of the microwave reflection coefficient due to the electromagnetic interaction between the resonator and the glucose solution. The in vitro results show the measured signal-to-noise ratio of about 34 dB, and the minimum detectible signal level of about 0.022 dB/(mg/dl). Our proposed system provides a unique approach for non-invasive and non-contact glucose monitoring, and it may serve as a bloodless glucometer.
