Between-grader repeatability of tear meniscus measurements using Fourier-domain OCT in patients with dry eye
Ethan H Tittler1, Matthew C Bujak, Pho Nguyen
1Keck School of Medicine, University of Southern California, Los Angeles, California, USA.
Background And Objective:
To examine the between-grader repeatability of height, depth, and cross-sectional area measurements of the lower tear meniscus, using a Fourier-domain optical coherence tomography (OCT) system.
Patients And Methods:
A total of 16 patients with dry eye had the lower tear meniscus of the right eye imaged twice in rapid succession. The tear meniscus height, depth, and cross-sectional area were measured by two masked graders using computer calipers. The between-grader variability, calculated using the pooled coefficient of variation (CV%), assessed the repeatability of the measurements.
Results:
The between-grader CV% was 12.1%, 15.7%, and 19.5% for height, depth, and area, respectively. The between-image variability was 17.1%, 13.4%, and 35.4% for height, depth, and area, respectively. The overall intraclass correlation was 99%. There was no systematic bias between the two graders.
Conclusion:
Fourier-domain OCT demonstrates good between-grader and between-image repeatability in measuring the height, depth, and cross-sectional area of the tear meniscus in patients with dry eye. Measurement variability was primarily due to the difference between images rather than graders.