Measuring and Modeling Contractile Drying in Human Stratum Corneum
Published on: March 1, 2017
Closed-cell foam skin thickness measurement using a scanning electron microscope
Clifford S Todd1, Valentina Kuznetsova
1The Dow Chemical Company, Analytical Sciences, Midland, MI 48667, USA. CTodd2@Dow.com
This study introduces a scanning electron microscope (SEM) method to measure polymer foam skin thickness using backscatter electron (BSE) imaging. Varying accelerating voltages allows for detailed characterization of skin thickness across the foam surface.
Area of Science:
- Materials Science
- Polymer Science
- Surface Characterization
Background:
- Closed cell polymer foams are widely used in various industries.
- Accurate characterization of foam skin thickness is crucial for performance.
- Existing methods for skin thickness measurement can be limited in scope or resolution.
Purpose of the Study:
- To develop and validate a non-destructive method for assessing closed cell polymer foam skin thickness.
- To demonstrate the utility of scanning electron microscopy (SEM) with backscatter electron (BSE) imaging for this purpose.
- To establish a technique applicable across a range of SEM instruments and experimental conditions.
Main Methods:
- Utilizing SEM to acquire BSE images of polymer foam samples at varying accelerating voltages (5 keV to 30 keV).
- Employing Monte Carlo modeling to correlate electron beam penetration depth with observed image contrast and skin thickness.
- Applying image analysis techniques for semiquantitative assessment of skin thickness distribution over large surface areas.
Main Results:
- Demonstrated that BSE imaging in SEM can effectively differentiate polymer skin thickness based on image contrast.
- Established a relationship between accelerating voltage and the detectable skin thickness, with higher voltages enabling measurement of thicker skins (up to ~4 μm).
- Showcased the method's applicability to real foam samples, allowing for qualitative and semiquantitative skin thickness assessments.
Conclusions:
- The proposed SEM-based BSE imaging technique provides a versatile and effective approach for characterizing polymer foam skin thickness.
- This method is accessible with standard SEM equipment capable of variable accelerating voltages and low-vacuum imaging.
- The technique offers a sensitive means to assess skin thickness distribution, crucial for optimizing foam properties and performance.