Monte Carlo simulation of contrast-enhanced whole brain radiotherapy on a CT scanner
Bernd-Joachim Schultz1, Peter Wust, Lutz Lüdemann
1Department of Radiotherapy, Charité Medical Center, 13353 Berlin, Germany.
Purpose:
To perform a feasibility study of contrast-enhanced whole brain radiotherapy for treating patients with multiple brain metastasis using a conventional computed tomography (CT) scanner.
Methods:
The treatment dose was optimized to be applied in a single run using a maximum tube power of 5200 kWs at 140 kV. CT scans of a large and a small head were used as reference. Irradiation geometry, shielding, axial beam collimation, radial beam collimation, gantry tilt, and tube current for beam modulation were optimized using a Monte Carlo simulation and a contrast agent concentration of 5 mg/ml iodine in the tumor. The statistical uncertainty of the Monte Carlo simulation was corrected using back convolution.
Results:
Using a CT tube with a beam collimation of 28.8 mm, a mean tumor dose of 1.76 +/- 0.13 Gy was achieved, while the head bone dose was 2.61 +/- 0.18 Gy with a normal brain dose of 0.98 +/- 0.06 Gy, eye dose of 0.19 +/- 0.05 Gy, and lens dose of 0.15 +/- 0.03 Gy, respectively. Using a CT tube with dose modulation and a beam collimation of 40.0 mm, the mean tumor dose was 2.00 +/- 0.11 Gy with a head bone dose of 1.96 +/- 0.14 Gy, normal brain dose of 1.13 +/- 0.08 Gy, eye dose of 0.21 +/- 0.05 Gy, and lens dose of 0.16 +/- 0.02 Gy, respectively. Thus a standard CT scanner enables an effective tumor dose of 37.0 Gy to be administered in 13 fractions, while exposing healthy brain to an effective dose of 17.2 Gy and head bone to 69.3 Gy. Additional radial collimation implemented in the hardware improves the therapeutic tumor dose by 25.2% in relation to the bone dose.
Conclusions:
Contrast-enhanced total brain radiotherapy is feasible using a conventional CT tube with optimized dose application.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
08:54Treatment of Liver Metastases Using an Internal Target Volume Method for Stereotactic Body Radiotherapy
Published on: May 8, 2018
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...