Image quality and radiation exposure using different low-dose scan protocols in dual-source CT coronary angiography:
Lisan A Neefjes1, Anoeshka S Dharampal, Alexia Rossi
1Department of Radiology, Erasmus Medical Center, Dr Molewaterplein 40, Room Hs 207, PO Box 2040, 3000 CA Rotterdam, the Netherlands.
High-pitch spiral CT coronary angiography is best for low heart rates (<55 bpm), offering lower radiation dose and comparable image quality. For higher heart rates, sequential scanning is preferred for optimal CT coronary angiography results.
Area of Science:
- Radiology
- Cardiovascular Imaging
- Medical Imaging Technology
Background:
- Computed tomographic (CT) coronary angiography is crucial for diagnosing coronary artery disease.
- Image quality and radiation dose are key considerations in CT coronary angiography protocols.
- Patient heart rate significantly influences the success and safety of CT coronary angiography.
Purpose of the Study:
- To compare image quality, radiation dose, and their relationship with heart rate for different CT coronary angiographic scan protocols.
- To evaluate the efficacy of high-pitch spiral, sequential (narrow and wide-window), and retrospective spiral scanning techniques.
- To determine optimal CT coronary angiography protocols based on patient heart rate using a 128-section dual-source CT scanner.
Main Methods:
- A study involving 272 patients undergoing CT coronary angiography was conducted.
- Patients were categorized into two groups based on heart rate: <65 bpm (Group A) and ≥65 bpm (Group B).
- Group A patients were randomized to high-pitch spiral or narrow-window sequential scanning; Group B patients received wide-window sequential or retrospective spiral scanning.
Main Results:
- In Group A, high-pitch spiral scanning yielded significantly lower image quality compared to sequential scanning (P < .001), except in those with heart rates <55 bpm.
- Image quality was comparable between sequential and retrospective spiral scanning in Group B (P = .54).
- Mean radiation doses were significantly lower with high-pitch spiral scanning in Group A and sequential scanning in Group B compared to alternative protocols within each group (P < .001 and P = .02, respectively).
Conclusions:
- High-pitch spiral CT coronary angiography protocols are recommended for patients with regular, low heart rates (<55 bpm) due to reduced radiation exposure.
- Sequential CT coronary angiography protocols are preferred for patients with heart rates of 65 bpm or higher.
- Protocol selection should be tailored to patient heart rate to optimize the balance between image quality and radiation dose in CT coronary angiography.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies III: Computed Tomography
Radiological Investigation I: X-ray and CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...