Screw locking elements: a means to modify the flexibility of osteoporotic fracture fixation with DCPs without
A Yánez1, A Cuadrado, J A Carta
1Department of Mechanical Engineering, Biomechanical Laboratory, Las Palmas University, Las Palmas de Gran Canaria, Spain.
Abstract:
This paper analyses whether it is possible to use dynamic compression plates (DCPs) and screw locking elements (SLEs) to vary the flexibility of osteoporotic fracture fixation without compromising the strength and stability of the construct. Compression, torsion and four-point bending static strength tests were conducted. Cyclic load tests of up to 10,000 load cycles were also carried out to determine stiffness performance. Four fixation systems were mounted onto polyurethane bone models. Group 1 consists of the DCP and six cortical screws. Group 2, idem, but with the addition of two SLEs. Group 3, idem, but with the addition of six SLEs. Group 4 used the locking compression plate (LCP) and locking screws. The results indicated no significant difference (p>0.05) in the strength of groups 2-4. It was also observed that the torsional stiffness of group 3 (0.30 Nm/°) was higher than that of group 2 (0.23 Nm/°) and similar to that of group 4 (0.28 Nm/°). Compression stiffness of group 4 (124 N/mm) was higher than that of group 2 (102 N/mm), but lower than that of group 3 (150 N/mm). No notable differences were observed for structural bending stiffness. It is concluded that by using the DCP with SLEs it is possible to modify the stiffness of the fixation construct for the repair of osteoporotic fractures and, in this way, facilitate the conditions suitable on secondary bone healing.
More Related Videos
10:09Adjustable Stiffness, External Fixator for the Rat Femur Osteotomy and Segmental Bone Defect Models
Published on: October 9, 2014
15:11Surgical Fixation of Sternal Fractures: Preoperative Planning and a Safe Surgical Technique Using Locked Titanium Plates and Depth Limited Drilling
Published on: January 5, 2015
Related Concept Videos
Frictional Forces on Screws
A jack with a square-threaded screw is a mechanical device used to lift heavy loads by applying a force at its handle. When the force is applied, the screw turns, raising the load. The screw can...
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.
Fractures: Bone Repair
Minor fractures with no bone displacement are treated by immobilizing the fractured bone using a cast or splint. However, in the case of fractures with displaced bones, the broken bones are repositioned before immobilization to ensure successful healing without deformation and loss of function. The realignment of fractured bone ends is performed through a process called reduction. If the procedure...