Related Experiment Videos
Patterns of failure in the original Besser bipolar prosthesis
1Department of Orthopaedic Surgery, University of Cape Town.
Summary
The Besser endoprosthesis showed a 50% failure rate in young patients with osteonecrosis of the femoral head, with issues like dislocation and component disassembly. This high failure rate is concerning for hip replacement outcomes.
Area of Science:
- Orthopedic surgery
- Biomedical engineering
- Hip reconstruction
Background:
- Osteonecrosis of the femoral head (ONFH) is a debilitating condition often affecting young patients.
- Total hip arthroplasty (THA) is a common treatment for advanced ONFH.
- The Besser endoprosthesis is a specific implant used in THA.
Purpose of the Study:
- To evaluate the clinical outcomes and failure modes of the original Besser endoprosthesis in young patients with ONFH.
- To identify factors contributing to the failure of this specific hip implant.
Main Methods:
- Retrospective review of 14 original Besser endoprostheses implanted in young patients with ONFH.
- Analysis of patient records, radiographic images, and surgical notes.
- Assessment of implant survival, dislocation rates, component failure, and patient-reported pain and mobility.
Main Results:
- A high failure rate of 50% was observed in this series.
- Complications included early cup dislocation (2 hips), acetabular degeneration requiring revision (1 hip), and component disassembly (4 hips).
- Persistent vertical cup orientation was associated with component failure and was present in 3 asymptomatic hips, indicating a predisposition to future failure.
Conclusions:
- The original Besser endoprosthesis demonstrated a disappointing failure rate in this cohort of young ONFH patients.
- Component disassembly and cup failure appear linked to a persistently vertical cup orientation.
- Further investigation and caution are warranted when considering this implant for hip reconstruction.