The effect of pulling out cochlear implant electrodes on inner ear microstructures: a temporal bone study.
Ingo Todt1, Rainer O Seidl, Arne Ernst
1Department of Otolaryngology, Unfallkrankenhaus Berlin, 12683 Berlin, Germany.
International Journal of Otolaryngology
|October 19, 2011
Summary
Cochlear implant electrode removal rarely damages cochlear structures. Careful removal, guided by forceps, prevents electrode tip turning, protecting the delicate basilar membrane.
Area of Science:
- Otoacoustic Emissions
- Neuroscience
- Biomedical Engineering
Background:
- Cochlear implant (CI) electrode revision surgeries are increasingly common.
- Microstructural effects of CI electrode extraction on the cochlea are not well understood.
Purpose of the Study:
- To investigate the microstructural consequences of cochlear implant electrode pull-out.
Main Methods:
- Histological analysis of 10 human temporal bones after CI electrode extraction.
- Video-capturing analysis of intrascalar electrode movements during pull-out in 9 temporal bones.
Main Results:
- No histological evidence of modiolar wall or spiral osseous lamina disruption.
- Basilar membrane elevation observed in one case, not definitively linked to electrode pull-out.
- Video analysis revealed electrode rotation in one instance, potentially elevating the basilar membrane.
Conclusions:
- Perimodiolar CI electrode extraction does not appear to cause microstructural damage to the modiolar wall.
- Guided removal, e.g., with forceps, is recommended to prevent electrode tip rotation towards the basilar membrane.