Related Experiment Video
Updated: May 28, 2026

Infrared Degenerate Four-wave Mixing with Upconversion Detection for Quantitative Gas Sensing
Published on: March 22, 2019
[Research on trace gas spectral measurement on intra-cavity fiber optic laser]
Hong-Xia Zhang1, Kun Liu, Da-Gong Jia
1College of Precision Instrument & Opto-electronics Engineering, Tianjin University, Tianjin 300072, China. hxzhang@tju.edu.cn
Abstract:
Due to the advantages of immunity of electrical/electronic, high performance cost ratio, remote detection and multiplexing capability, intra-cavity fiber optic gas measurement has aroused wide concern. The trace gas measurement system has been developed based on the elaborated gas cell and reflector. The wavelength sweeping technique (WST) is realized when the Fabry-Perot type tunable optical filter is applied by the sawtooth driver voltage. Multi absorption lines can be obtained and one scanning measurement with WST is equal to multiple independent detections, so the gas measurement sensitivity is improved remarkably. The experimental results show that the acetylene detection sensitivity is reduced to less than 100 ppm and the relative measurement error is less than 3% of practical gas concentration.